National Instruments Graphics Tablet NI PCI 5114 User Manual

NI PXI/PCI-5114 Specifications  
8-Bit 250 MS/s Digitizer  
This document lists the specifications for the NI PXI/PCI-5114 (NI 5114)  
high-speed digitizer. Unless otherwise noted, these specifications are valid  
for the following conditions:  
All filter settings  
All impedance selections  
Sample clock set to 250 MS/s  
Typical values are representative of an average unit operating at room  
temperature. Specifications are subject to change without notice. For the  
most recent NI 5114 specifications, visit ni.com/manuals.  
To access the NI 5114 documentation, including the NI High-Speed  
Digitizers Getting Started Guide, which contains functional descriptions  
of the NI 5114 signals, navigate to Start»All Programs»National  
Instruments»NI-SCOPE»Documentation.  
cause burns. Allow the NI 5114 to cool before removing it from the PXI chassis or PC.  
Phase-Locked Loop (PLL) Reference Clock................................... 7  
Trigger..................................................................................................... 8  
Reference (Stop) Trigger ................................................................. 8  
TRIG (External Trigger, Front Panel Connector)............................ 10  
PFI 0 and PFI 1 (Programmable Function Interface,  
AUX Front Panel Connectors)............................................................. 11  
TClk Specifications................................................................................. 12  
 
Specification  
Value  
Comments  
Voltage Levels  
Full Scale (FS)  
InputRangeand  
Programmable  
Vertical Offset  
50 Ω  
Vertical Offset  
1 MΩ  
Vertical Offset  
Range  
Range  
(Vpk-pk  
0.04  
0.1  
0.2  
0.4  
1
)
Range (V)  
(Vpk-pk  
0.04  
0.1  
0.2  
0.4  
1.0  
2.0  
4.0  
10  
)
Range (V)  
0.8  
0.8  
0.8  
0.8  
0.8  
0.8  
0.8  
0.8  
6.5  
8.0  
2
6.0  
8.0  
4
5.0  
8.0  
30  
10  
2.0  
20  
25  
40  
15  
MaximumInput  
Overload  
50 Ω  
1 MΩ  
|Peaks| 35 V  
7 Vrms with |Peaks| 10 V  
Accuracy  
Resolution  
8 bits  
DC Accuracy  
(Programmable  
Vertical  
NI PXI-5114:  
(1.5% of Input + 0.3% of FS + 200 μV)  
Within 5 °C of  
self-calibration  
temperature  
NI PCI-5114:  
(1.5% of Input + 0.3% of FS + 280 μV)  
Offset = 0 V)  
Programmable  
Vertical Offset  
Accuracy  
2% of offset setting  
Within 5 °C of  
self-calibration  
temperature  
DC Drift  
(0.03% of Input + 0.06% of FS + 40 μV) per °C  
–60 dB at 10 MHz  
Crosstalk,  
Typical  
CH 0 to/from  
CH 1, External  
Trigger to CH 0  
or CH 1  
–45 dB at 100 MHz  
© National Instruments Corporation  
3
NI PXI/PCI-5114 Specifications  
 
Specification  
Value  
Comments  
Bandwidth and Transient Response  
Bandwidth  
(–3 dB)  
Range  
Rise/FallTime,  
Typical  
(Vpk-pk  
)
Bandwidth  
125 MHz  
All ranges  
2.8 ns  
except 0.04  
0.04  
100 MHz  
3.5 ns  
Bandwidth  
Limit Filter  
20 MHz Noise Filter  
AC Coupling*  
Cutoff (–3 dB),  
Typical  
*AC coupling  
available on  
1 MΩ only  
12 Hz  
Passband  
Flatness  
Referenced to  
50 kHz  
1 dB up to 50 MHz  
Bandwidth  
limit filter off  
1
0
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
Full Bandwidth  
Noise Filter  
–10  
1 M  
10 M  
100 M  
200 M  
Frequency (Hz)  
Figure 1. NI 5114 Frequency Response (Typical)  
NI PXI/PCI-5114 Specifications  
4
ni.com  
 
Specification  
Value  
Comments  
Spectral Characteristics  
Spurious-Free  
Range (Vpk-pk  
)
10 MHz,  
Dynamic Range  
with Harmonics  
(SFDR), Typical  
–1 dBFS  
input signal  
All ranges except 0.04  
0.04  
58 dBc  
58 dBc  
Includes the  
2nd through  
the 5th  
Total Harmonic  
Distortion  
(THD), Typical  
–58 dBc  
–58 dBc  
6.2  
harmonics  
Measured  
from DC to  
125 MHz  
Effective  
Number of Bits  
(ENOB),  
7.2  
Calculated*  
20 MHz  
bandwidth  
limit filter off  
Signal to Noise  
and Distortion  
(SINAD),  
44 dB  
38 dB  
Typical  
RMS Noise  
Range  
(Vpk-pk  
20 MHz  
Filter On  
20 MHz  
Filter Off  
50 Ω  
)
terminator  
connected  
to input  
All ranges  
except 0.04  
0.28% FS  
0.28% FS  
0.28% FS  
0.04  
0.45% FS  
*ENOB = log2(sinad) – ½log2(1.5) – log2(A/V)  
where  
sinad = the linear representation of SINAD  
A = amplitude of the supplied sine wave during the test  
V = (peak) full-scale range of the waveform recorder input  
Refer to 1057-1994 IEEE Standard for Digitizing Waveform Recorders for information on equation derivation.  
© National Instruments Corporation  
5
NI PXI/PCI-5114 Specifications  
 
Horizontal  
Sample Clock  
Specification  
Value  
Comments  
Sources  
Internal, Onboard Clock (internal VCXO)*  
External, CLK IN (front panel SMB connector)  
* Internal Sample  
Clock is locked to the  
Reference Clock or  
derived from the  
onboard VCXO  
Onboard Clock (Internal VCXO)  
Sample Rate  
Range  
Real-Time Sampling  
(Single Shot)  
Random Interleaved  
Sampling (RIS)  
* Divide by n  
decimation used  
for all rates less than  
250 MS/s  
3.815 kS/s to 250MS/s* 250 MS/s to 5 GS/s in  
increments of 250 MS/s  
Formoreinformation  
about Sample Clock  
and decimation, refer  
to the NI High-Speed  
Digitizers Help.  
Timebase  
Frequency  
250 MHz  
When not using  
External Sample  
Clock  
Timebase  
Accuracy  
Not Phase-Locked to  
Reference Clock  
Phase-Locked to  
Reference Clock  
25 ppm  
Equal to the Reference  
Clock accuracy  
Sample Clock  
Delay Range  
1 Sample Clock period  
20 ps  
Sample Clock  
Delay Resolution  
NI PXI/PCI-5114 Specifications  
6
ni.com  
 
   
Specification  
Value  
Comments  
External Sample Clock  
Sources  
CLK IN (front panel SMB connector)  
50 MHz to 250 MHz  
Frequency  
Range  
Divide by n  
decimation  
available where  
1 n 65,535  
Formoreinformation  
about Sample Clock  
and decimation, refer  
to the NI High-Speed  
Digitizers Help.  
Duty Cycle  
Tolerance  
45% to 55%  
Phase-Locked Loop (PLL) Reference Clock  
Specification  
Value  
Sources  
NI PXI-5114  
PXI_CLK10  
NI PCI-5114  
RTSI 7  
CLK IN  
(front panel SMB connector)  
(backplane connector)  
CLK IN  
(front panel SMB connector)  
Frequency  
Range  
1 MHz to 20 MHz in 1 MHz increments  
Default of 10 MHz  
The PLL Reference Clock frequency must be accurate to 50 ppm  
45% to 55%  
Duty Cycle Tolerance  
Exported Reference  
Clock Destinations  
NI PXI-5114  
NI PCI-5114  
PFI <0..1> (front panel 9-pin  
mini-circular DIN connector)  
PFI <0..1> (front panel 9-pin  
mini-circular DIN connector)  
PXI_Trig <0..7> (backplane  
connector)  
RTSI <0..7>  
© National Instruments Corporation  
7
NI PXI/PCI-5114 Specifications  
 
 
CLK IN (Sample Clock and Reference Clock Input,  
Front Panel Connector)  
Specification  
Value  
Input Voltage Range  
Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)  
Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk  
Maximum Input  
Overload  
7 Vrms with |Peaks| 10 V  
Impedance  
Coupling  
50 Ω  
AC  
Trigger  
Reference (Stop) Trigger  
Specification  
Value  
Comments  
Trigger Types  
and Sources  
Types  
Sources  
Refer to the  
following  
sections and to  
NI High-Speed  
Digitizers Help  
for more  
Edge, Window, Hysteresis,  
Video, Digital, Immediate,  
and Software  
CH 0, CH 1, TRIG,  
PXI_Trig<0..6>, PFI <0..1>,  
PXI Star Trigger,  
RTSI<0..6>, and Software  
information.  
Time  
Resolution  
TDC  
On  
Onboard Clock  
External Clock  
N/A  
TDC = Time to  
Digital  
Conversion  
Circuit  
40 ps  
4 ns  
Off  
External Clock Period  
Rearm Time  
10 μs  
Rearm Time  
TDC  
On  
Holdoff set to 0  
Off  
2 μs  
Holdoff  
From Rearm Time up to [(235 – 1) × (Sample Clock Period)]  
Trigger Delay  
From 0 up to [(235 – 1) – posttrigger samples] ×  
(1/sample rate), in seconds  
NI PXI/PCI-5114 Specifications  
8
ni.com  
 
     
Specification  
Value  
Comments  
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)  
Sources  
CH 0 (front panel BNC connector)  
CH 1 (front panel BNC connector)  
TRIG (front panel BNC connector)  
8 bits (1 in 256)  
Trigger Level  
Resolution  
Trigger Level  
Range  
CH 0, CH 1  
100% FS  
TRIG (External Trigger)  
5 V  
Edge Trigger  
Sensitivity  
5% FS up to 100 MHz  
0.5 Vpk-pk up to 100 MHz  
Level  
5% FS up to 10 MHz  
0.5 V up to 10 MHz  
Accuracy,  
Typical  
Jitter  
65 ps rms  
Trigger Filters Low Frequency (LF) Reject High Frequency (HF) Reject  
50 kHz  
50 kHz  
Digital Trigger (Digital Trigger Type)  
Sources  
NI PXI-5114  
NI PCI-5114  
PXI_Trig <0..6> (backplane RTSI <0..6>  
connector)  
PFI <0..1> (front panel  
PFI <0..1> (front panel  
SMB connector)  
SMB connector)  
PXI Star Trigger (backplane  
connector)  
Video Trigger (Video Trigger Type)  
Sources CH 0 (front panel BNC connector)  
CH 1 (front panel BNC connector)  
TRIG (front panel BNC connector)  
© National Instruments Corporation  
9
NI PXI/PCI-5114 Specifications  
 
Specification  
Value  
Comments  
Types  
Specific Line  
Any Line  
Specific Field  
Standards  
SDTV: M-NTSC, B/G-PAL, SECAM, M-PAL  
fps = fields per  
second  
EDTV: 480i/59.94 fps, 480i/60 fps, 480p/59.94 Fps,  
480p/60 Fps, 576i/50 fps, 576p/50 Fps  
Fps = Frames  
per second  
HDTV: 720p/50 Fps, 720p/59.94 Fps, 720p/60 Fps,  
1080i/50 fps, 1080i/59.94 fps, 1080i/60 fps, 1080p/24 Fps  
TRIG (External Trigger, Front Panel Connector)  
Specification  
Value  
Connector  
BNC  
Impedance  
1 MΩ in parallel with 22 pF  
Coupling  
AC, DC  
12 Hz  
AC-Coupling Cutoff (–3 dB)  
Input Voltage Range  
5 V  
Maximum Input Overload  
|Peaks| 42 V  
NI PXI/PCI-5114 Specifications  
10  
ni.com  
 
 
PFI 0 and PFI 1 (Programmable Function Interface,  
AUX Front Panel Connectors)  
Specification  
Connector  
Value  
9-pin mini-circular DIN  
Bi-directional  
Direction  
As an Input (Trigger)  
Destinations  
Start Trigger (Acquisition Arm)  
Reference (Stop) Trigger  
Arm Reference Trigger  
Advance Trigger  
150 kΩ  
Input Impedance  
VIH  
2.0 V  
VIL  
0.8 V  
Maximum Input Overload  
–0.5 V, 5.5 V  
Maximum Frequency  
25 MHz  
As an Output (Event)  
Sources  
Start Trigger (Acquisition Arm)  
Reference (Stop) Trigger  
End of Record  
Done (End of Acquisition)  
Probe Compensation (1 kHz, 50% duty cycle square wave,  
PFI 1 only)  
Output Impedance  
Logic Type  
50 Ω  
3.3 V CMOS  
24 mA  
Maximum Drive Current  
Maximum Frequency  
25 MHz  
© National Instruments Corporation  
11  
NI PXI/PCI-5114 Specifications  
 
 
TClk Specifications  
National Instruments TClk synchronization method and the NI-TClk driver  
are used to align the sample clocks on any number of SMC-based modules  
in a chassis. For more information about TClk synchronization, refer to the  
NI-TClk Synchronization Help, which is located within the NI High-Speed  
Digitizers Help.  
Specifications are valid for any number of modules installed in one  
NI PXI-1042 chassis.  
All parameters set to identical values for each SMC-based module.  
Sample Clock set to 250 MS/s and all filters are disabled.  
For other configurations, including multichassis systems, contact  
NI Technical Support at ni.com/support.  
Note Although you can use NI-TClk to synchronize nonidentical modules, these  
specifications apply only to synchronizing identical modules.  
Specification  
Value  
Comments  
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)  
Skew  
500 ps  
Caused by clock and analog  
path delay differences  
No manual adjustment  
performed  
Skew After Manual <20 ps  
Adjustment  
For information about  
manual adjustment, refer  
to the Synchronization  
Repeatability Optimization  
topic in the NI-TClk  
Synchronization Help.  
For additional help with the  
adjustment process, contact  
NI Technical Support at  
ni.com/support.  
Sample Clock  
Adjustment  
Resolution  
<20 ps  
NI PXI/PCI-5114 Specifications  
12  
ni.com  
 
 
Waveform Specifications  
Specification  
Value  
8 megasamples  
Comments  
Onboard Memory  
Size  
8 MB per Channel  
Standard  
per channel  
64 MB per Channel  
Option  
64 megasamples  
per channel  
256 MB per Channel  
Option  
256 megasamples  
per channel  
Minimum Record  
Length  
1 Sample  
Number of  
Pretrigger Samples  
Zero up to full Record Length  
Zero up to full Record Length  
Single-record  
mode and  
multiple-record  
mode  
Number of  
Posttrigger Samples  
Single-record  
mode and  
multiple-record  
mode  
Maximum Number  
of Records in  
Onboard Memory  
8 MB/channel  
64 MB/channel  
256 MB/channel  
32,768  
* It is possible  
to exceed these  
numbers if you  
fetch records while  
acquiring data. For  
more information,  
refer to the  
100,000*  
100,000*  
NI High-Speed  
Digitizers Help.  
Allocated Onboard  
Memory per Record  
(Record Length × 1 byte/S) + 240 bytes,  
rounded up to next multiple of 128 bytes  
or  
256 bytes, whichever is greater  
© National Instruments Corporation  
13  
NI PXI/PCI-5114 Specifications  
 
 
Calibration  
Specification  
Value  
Self-Calibration  
Self-calibration is done on software command. The calibration  
corrects for gain, offset, compensated 1 MΩ attenuator, triggering,  
and timing adjustment errors for all input ranges.  
External Calibration  
(Factory Calibration)  
The external calibration calibrates the VCXO, gain, and the voltage  
reference. Appropriate constants are stored in nonvolatile memory.  
Interval for External  
Calibration  
2 years  
Warm-Up Time  
15 minutes  
Power  
Specification  
Typical Value  
NI PXI-5114  
840 mA  
1.1 A  
NI PCI-5114  
1.6 A  
+3.3 VDC  
+5 VDC  
1.7 A  
+12 VDC  
–12 VDC  
Total Power  
250 mA  
170 mA  
13.32 W  
45 mA  
14.32 W  
NI PXI/PCI-5114 Specifications  
14  
ni.com  
 
   
Software  
Specification  
Value  
NI PXI-5114: NI-SCOPE 2.9 or later  
Driver Software  
NI PCI-5114: NI-SCOPE 3.1 or later  
NI-SCOPE is an IVI-compliant driver that allows you to configure,  
control, and calibrate the NI 5114. NI-SCOPE provides application  
programming interfaces for many development environments.  
Application Software  
NI-SCOPE provides programming interfaces, documentation, and  
examples for the following application development environments:  
LabVIEW  
LabWindows/CVI™  
Measurement Studio  
Microsoft Visual C/C++  
Microsoft Visual Basic  
Interactive Soft Front  
Panel and  
The Scope Soft Front Panel 2.3 or later supports interactive control of the  
NI 5114. The Scope Soft Front Panel is included on the NI-SCOPE CD.  
Configuration  
National Instruments Measurement & Automation Explorer (MAX) also  
provides interactive configuration and test tools for the NI 5114. MAX is  
included on the NI-SCOPE CD.  
© National Instruments Corporation  
15  
NI PXI/PCI-5114 Specifications  
 
 
Environment  
NI PXI-5114  
Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in  
the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit.  
The NI PXI-5114 is intended for indoor use only.  
Specification  
Value  
0 ºC to +55 ºC in all NI PXI chassis except the following:  
0 ºC to +45 ºC when installed in an NI PXI-1000/B or PXI-101x chassis  
Meets IEC-60068-2-1 and IEC-60068-2-2  
–40 ºC to +71 ºC  
Operating Temperature  
Storage Temperature  
Meets IEC-60068-2-1 and IEC-60068-2-2  
10% to 90%, noncondensing  
Operating Relative  
Humidity  
Meets IEC-60068-2-56  
Storage Relative  
Humidity  
5% to 95%, noncondensing  
Meets IEC-60068-2-56  
Operating Shock  
30 g, half-sine, 11 ms pulse  
Meets IEC-60068-2-27  
Test profile developed in accordance with MIL-PRF-28800F  
50 g, half-sine, 11 ms pulse  
Storage Shock  
Meets IEC-60068-2-27  
Test profile developed in accordance with MIL-PRF-28800F  
5 Hz to 500 Hz, 0.31 grms  
Operating Vibration  
Storage Vibration  
Meets IEC-60068-2-64  
5 Hz to 500 Hz, 2.46 grms  
Meets IEC-60068-2-64  
Test profile exceeds requirements of MIL-PRF-28800F, Class B  
2,000 m maximum (at 25 °C ambient temperature)  
2
Altitude  
Pollution Degree  
NI PXI/PCI-5114 Specifications  
16  
ni.com  
 
   
NI PCI-5114  
Note To ensure that the NI PCI-5114 cools effectively, make sure that the chassis in which  
it is used has active cooling that provides at least some airflow across the PCI card cage.  
To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.  
Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit  
for important cooling information. The NI PCI-5114 is intended for indoor use only.  
Specification  
Value  
Operating Temperature  
0 ºC to +45 ºC  
Meets IEC-60068-2-1 and IEC-60068-2-2  
–40 ºC to +71 ºC  
Storage Temperature  
Meets IEC-60068-2-1 and IEC-60068-2-2  
10% to 90%, noncondensing  
Operating Relative  
Humidity  
Meets IEC-60068-2-56  
Storage Relative  
Humidity  
5% to 95%, noncondensing  
Meets IEC-60068-2-56  
Storage Shock  
50 g, half-sine, 11 ms pulse  
Meets IEC-60068-2-27  
Test profile developed in accordance with MIL-PRF-28800F  
5 Hz to 500 Hz, 2.46 grms  
Storage Vibration  
Meets IEC-60068-2-64  
Test profile exceeds requirements of MIL-PRF-28800F, Class B  
2,000 m maximum (at 25 °C ambient temperature)  
2
Altitude  
Pollution Degree  
© National Instruments Corporation  
17  
NI PXI/PCI-5114 Specifications  
 
 
Safety, Electromagnetic Compatibility,  
and CE Compliance  
Safety  
This product is designed to meet the requirements of the following  
standards of safety for electrical equipment for measurement, control,  
and laboratory use:  
IEC 61010-1, EN 61010-1  
UL 61010-1, CSA 61010-1  
Note For UL and other safety certifications, refer to the product label or visit ni.com/  
certification, search by model number or product line, and click the appropriate link  
in the Certification column.  
Electromagnetic Compatibility  
This product is designed to meet the requirements of the following  
standards of EMC for electrical equipment for measurement, control, and  
laboratory use:  
EN 61326 EMC requirements; Minimum Immunity  
EN 55011 Emissions; Group 1, Class A  
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A  
Note For EMC compliance, operate this device with RG223/U or equivalent shielded  
cable. Operate according to product documentation  
CE Compliance  
This product meets the essential requirements of applicable European  
Directives, as amended for CE marking, as follows:  
2006/95/EC; Low-Voltage Directive (safety)  
2004/108/EC; Electromagnetic Compatibility Directive (EMC)  
Note Refer to the Declaration of Conformity (DoC) for this product for any additional  
regulatory compliance information. To obtain the DoC for this product, visit ni.com/  
certification, search by model number or product line, and click the appropriate link  
in the Certification column.  
NI PXI/PCI-5114 Specifications  
18  
ni.com  
 
       
Environmental Management  
National Instruments is committed to designing and manufacturing  
products in an environmentally responsible manner. NI recognizes that  
eliminating certain hazardous substances from our products is beneficial  
not only to the environment but also to NI customers.  
For additional environmental information, refer to the NI and the  
Environment Web page at ni.com/environment. This page contains the  
environmental regulations and directives with which NI complies, as well  
as any other environmental information not included in this document.  
Waste Electrical and Electronic Equipment (WEEE)  
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling  
center. For more information about WEEE recycling centers and National Instruments  
WEEE initiatives, visit ni.com/environment/weee.htm.  
⬉ᄤֵᙃѻક∵ᶧ᥻
ㅵ⧚
⊩ ˄Ё೑  
˅
RoHS  
Ё೑ᅶ᠋  
݇Ѣ  
ヺড়Ё೑⬉ᄤֵᙃѻકЁ䰤
Փ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ  
DŽ
National Instruments  
National Instruments  
(RoHS)  
Ё೑  
ড়㾘ᗻֵᙃˈ䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ  
RoHS  
ni.com/environment/rohs_china  
(For information about China RoHS compliance, go to  
.)  
Physical  
Front Panel Connectors  
Label  
CH 0  
Function  
Connector Type  
BNC female  
Comments  
Analog Input  
Analog Input  
External Trigger  
CH 1  
BNC female  
BNC female  
SMB jack  
TRIG  
CLK IN  
Sample Clock Input and  
Reference Clock Input  
AUX I/O  
PFI 0, PFI 1  
9-pin mini-circular DIN  
© National Instruments Corporation  
19  
NI PXI/PCI-5114 Specifications  
 
     
NI PXI-5114 Front Panel Indicators  
Label  
Function  
For more  
information,  
refer to the  
NI High-Speed  
Digitizers Help.  
ACCESS  
The ACCESS LED indicates the status of the PCI bus and  
the interface from the NI PXI-5114 to the controller.  
ACTIVE  
The ACTIVE LED indicates the status of the onboard  
acquisition hardware of the NI PXI-5114.  
Dimensions and Weight  
NI PXI-5114  
Dimensions  
3U, One slot, PXI/cPCI Module  
21.6 × 2.0 × 13.0 cm  
(8.5 × 0.8 × 5.1 in.)  
2.0 cm  
(0.8 in.)  
13.0 cm  
(5.1 in.)  
21.6 cm  
(8.5 in.)  
Weight  
455 g (16 oz)  
NI PXI/PCI-5114 Specifications  
20  
ni.com  
 
 
NI PCI-5114  
Dimensions  
35.5 × 2.0 × 11.3 cm  
(14.0 × 0.8 × 4.4 in.)  
2.0 cm  
(0.8 in.)  
11.3 cm  
(4.4 in.)  
35.5 cm  
(14.0 in.)  
Weight  
421 g (14.8 oz)  
© National Instruments Corporation  
21  
NI PXI/PCI-5114 Specifications  
 
Where to Go for Support  
The National Instruments Web site is your complete resource for technical  
support. At ni.com/support you have access to everything from  
troubleshooting and application development self-help resources to email  
and phone assistance from NI Application Engineers.  
A Declaration of Conformity (DoC) is our claim of compliance with the  
Council of the European Communities using the manufacturer’s  
declaration of conformity. This system affords the user protection for  
electronic compatibility (EMC) and product safety. You can obtain the DoC  
for your product by visiting ni.com/certification. If your product  
supports calibration, you can obtain the calibration certificate for your  
product at ni.com/calibration.  
National Instruments corporate headquarters is located at  
11500 North Mopac Expressway, Austin, Texas, 78759-3504.  
National Instruments also has offices located around the world to help  
address your support needs. For telephone support in the United States,  
create your service request at ni.com/support and follow the calling  
instructions or dial 512 795 8248. For telephone support outside the United  
States, contact your local branch office:  
Australia 1800 300 800, Austria 43 662 457990-0,  
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,  
Canada 800 433 3488, China 86 21 5050 9800,  
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,  
Finland 358 (0) 9 725 72511, France 01 57 66 24 24,  
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,  
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400,  
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,  
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,  
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,  
Poland 48 22 3390150,Portugal 351 210 311 210,Russia 7 495 783 6851,  
Singapore 1800 226 5886, Slovenia 386 3 425 42 00,  
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,  
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,  
Taiwan 886 02 2377 2222, Thailand 662 278 6777,  
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545  
NI PXI/PCI-5114 Specifications  
22  
ni.com  
 
 
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.  
Refer to the Terms of Use section on ni.com/legal for more information about National  
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade  
names of their respective companies. For patents covering National Instruments products, refer to the  
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or  
ni.com/patents.  
© 2005–2007 National Instruments Corporation. All rights reserved.  
374179C-01  
Aug07  
 

Miele Refrigerator KF 9757 iD User Manual
Mr Handsfree GPS Receiver Blue Eagle II User Manual
MTD Snow Blower 769 01275D User Manual
MTD Snow Blower 769 04101 User Manual
Nexus 21 TV Antenna NX2600 User Manual
Oki Copier ES3640E User Manual
Onkyo Stereo Receiver TX NR905 User Manual
Oster Cookware BYST EK18 User Manual
Paradyne Network Card Adapter Bracket User Manual
Parrot GPS Receiver SPV C600 User Manual